| Ireferensi Ejwayelekile | Izihloko |
| I-EN 60601-2-54:2009 | Imishini kagesi yezokwelapha - Ingxenye 2-54: Izidingo ezikhethekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile kwemishini ye-X-ray ye-radiography kanye ne-radioscopy |
| IEC60526 | Ipulaki yekhebula ene-voltage ephezulu kanye nokuxhumeka kwesokhethi kwemishini ye-X-ray yezokwelapha |
| I-IEC 60522:1999 | Ukunqunywa kokuhlunga okungapheli kwemihlangano ye-X-ray tube |
| I-IEC 60613-2010 | Izici zikagesi, ukushisa kanye nokulayisha kwamashubhu e-X-ray anode ajikelezayo ukuze kutholakale ukuxilongwa kwezokwelapha |
| IEC60601-1:2006 | Imishini kagesi yezokwelapha - Ingxenye 1: Izidingo ezijwayelekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile |
| I-IEC 60601-1-3:2008 | Imishini kagesi yezokwelapha - Ingxenye 1-3: Izidingo ezijwayelekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile - Izinga Lokubambisana: Ukuvikelwa kwemisebe emishinini ye-X-ray yokuxilonga |
| IEC60601-2-28:2010 | Imishini kagesi yezokwelapha - Ingxenye 2-28: Izidingo ezikhethekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile kwemihlangano yamashubhu e-X-ray yokuxilongwa kwezokwelapha |
| I-IEC 60336-2005 | Imishini kagesi yezokwelapha-imihlangano yamashubhu e-X-ray yokuxilongwa kwezokwelapha-Izimpawu zezindawo ezigxile kuzo |
●Igama lakhiwe kanje:
| MWHX7110A | Ipayipi | A | Isokhethi ye-voltage ephezulu enesiqondiso sama-degree angu-90 |
| MWTX71-0.6/1.2-125 | B | Isokhethi ye-voltage ephezulu enesiqondiso sama-degree angu-270 |
| Impahla | Imininingwane | Okujwayelekile | |
| Amandla okufaka ajwayelekile e-anode | F 1 | F 2 | I-IEC 60613 |
| 20kW(50/60Hz) | 40kW(50/60Hz) | ||
| Umthamo wokugcina ukushisa we-Anode | 110 kJ (150kHU) | I-IEC 60613 | |
| Umthamo omkhulu wokupholisa we-anode | 500W | ||
| Umthamo wokugcina ukushisa | 900kJ | ||
| Ukushisa okuqhubekayo okuphezulu ngaphandle komoya ojikelezayo | 180W | ||
| Izinto ze-anodeIzinto zokumboza eziphezulu ze-Anode | I-Rhenium-Tungsten-TZM(RTM) I-Rhenium-Tungsten-(RT) | ||
| I-engeli eqondiwe (Inkomba: i-axis yokubhekisela) | 12.5 ° | I-IEC 60788 | |
| Ukuhlanganiswa kwe-X-ray tube filtration engokwemvelo | 1.5 mm Al / 75kV | I-IEC 60601-1-3 | |
| Inani(amanani) lendawo yokugxila | F1 (ukugxila okuncane) | F2 (ukugxila okukhulu) | I-IEC 60336 |
| 0.6 | 1.2 | ||
| I-voltage ejwayelekile ye-X-ray tubeI-X-rayI-Fluoroscopic | 125kV 100kV | I-IEC 60613 | |
| Idatha ngokushisa kwe-cathode Umkhawulo wamanje I-voltage ephezulu | ≈ /AC, < 20 kHz | ||
| F1 | F 2 | ||
| 5.1A ≈7~9V | 5.1 A ≈12~14 V | ||
| Imisebe evuzayo ku-150 kV / 3mA ngebanga elingu-1m | ≤0.5mGy/h | IEC60601-1-3 | |
| Insimu yemisebe ephezulu | 443×443mm ku-SID 1m | ||
| Isisindo somhlangano we-X-ray tube | Cishe 18 kg | ||
| Imikhawulo | Imikhawulo Yokusebenza | Imikhawulo Yezokuthutha Neyokugcina |
| Izinga lokushisa elizungezile | Kusukela ku-10℃kuya ku-40℃ | Kusukela ku- 20℃to 70℃ |
| Umswakama ohlobene | ≤75% | ≤93% |
| Ingcindezi ye-barometric | Kusukela ku-70kPa kuya ku-106kPa | Kusukela ku-70kPa kuya ku-106kPa |
i-stator yesigaba esi-1
| Iphuzu lokuhlola | C-M | C-A |
| Ukumelana nokugoba | ≈18.0…22.0Ω | ≈45.0…55.0Ω |
| I-voltage yokusebenza evumelekile ephezulu (ukusebenza) | 230V±10% | |
| Ncoma i-voltage yokusebenza (ukusebenza) | 160V±10% | |
| I-voltage yokubhuleka | 70VDC | |
| I-voltage esebenzayo ekuchayekeni | Ama-Vrm angu-80 | |
| I-voltage egijimayo ku-fluoroscopy | Ama-VV angu-20-40 | |
| Isikhathi sokusebenza (kuye ngokuthi uhlelo lokuqalisa lunjani) | 1.2s | |
1. Imisebe ye-X-rayisivikelo
Lo mkhiqizo uhlangabezana nezidingo ze-IEC 60601-1-3.
Lo mhlangano we-X-ray tube ukhipha imisebe ye-X-ray isebenza. Ngakho-ke ngabasebenzi abafanelekayo nabaqeqeshwe ngokufanayo kuphela abavunyelwe ukusebenzisa umhlangano we-X-ray tube.
Imiphumela efanele yomzimba ingabangela ukulimala kwesiguli, abakhiqizi besistimu kufanele bavikeleke kahle ukuze bagweme imisebe ye-ionization.
2. I-Dielectric 0il
Ukuhlanganiswa kweshubhu le-X-ray kune-0il equkethwe yi-dielectric ukuze kuqiniswe ugesi ophezulu. Njengoba kuyingozi empilweni yabantu,uma ivezwe endaweni engavinjelwe,kufanele ilahlwe ngokulandela umthetho wendawo.
3.Umkhathi Wokusebenza
Ukuhlanganiswa kweshubhu le-X-ray akuvunyelwe ukusetshenziswa emoyeni wegesi evuthayo noma ebolile ·
4.Lungisa i-Tube Current
Kuye ngezimo zokusebenza,izici ze-filament zingase zishintshe.
Lolu shintsho lungase luholele ekuchayekeni ngokweqile kwe-X-ray tube assembly.
Ukuvimbela ukuthi i-X-ray tube assembly ingonakali,lungisa njalo ugesi wepayipi.
Ngaphandle kwalokho lapho ithubhu le-X-ray linenkinga yokugobalukusetshenziswa kwesikhathi eside,Kudingeka ukulungiswa komsinga wepayipi.
5Izinga lokushisa lezindlu ze-X-ray Tube
Ungathinti indawo engaphansi kwe-X-ray tube ngemuva nje kokuhlinzwa ngenxa yokushisa okuphezulu.
Ishubhu le-X-ray elihlala lipholile.
6.Imikhawulo yokusebenza
Ngaphambi kokusebenzisa,sicela uqinisekise ukuthi isimo semvelo singaphakathi kwe-Iimits yokusebenza.
7Noma yikuphi ukungasebenzi kahle
P1lula ukuxhumana ne-SAILRAY ngokushesha,uma kukhona ukungasebenzi kahle kwe-X-ray tube assembly okubonakalayo.
8. Ukulahlwa
Ukuhlanganiswa kweshubhu le-X-ray kanye neshubhu kuqukethe izinto ezifana namafutha nezinsimbi ezisindayo okumele kuqinisekiswe ukuthi ukulahlwa kwazo okungenangozi emvelweni nokufanelekile ngokuhambisana nemithethonqubo yezomthetho kazwelonke esebenzayo kumele kuqinisekiswe. Ukulahlwa kukadoti wasekhaya noma wezimboni akuvunyelwe. Umenzi unolwazi lobuchwepheshe oludingekayo futhi uzobuyisela ukuhlanganiswa kweshubhu le-X-ray ukuze kulahlwe.
Sicela uxhumane nesevisi yamakhasimende ngale njongo.
Uma(A) Indawo Encane Yokugxila
Uma(A) Indawo Enkulu Yokugxila
Izimo: I-Tube Voltage Isigaba Esithathu
Imvamisa Yamandla e-Stator 50Hz/60Hz
Ngiyi)
ama-t(s)
Ngiyi)
ama-t(s)
IEC60613
Izici Zokushisa Kwezindlu
SRMWHX7110A
Ukuhlanganiswa Kwesihlungi Nesigaba Esinqamulayo Sechweba
Izintambo Zesixhumi Se-Rotor
Inani Elincane Le-oda: 1pc
Intengo: Ingxoxo
Imininingwane Yokupakisha: 100pcs ngebhokisi ngalinye noma ngezifiso ngokuya ngobuningi
Isikhathi Sokulethwa: 1 ~ 2 amasonto ngokuya ngobuningi
Imigomo Yokukhokha: 100% T/T kusengaphambili noma i-WESTERN UNION
Amandla Okunikezela: 1000pcs/ngenyanga