Incazelo emfushane:

Isicelo: Ukuhlanganiswa kwe-X-ray tube kuzo zonke izivivinyo zokuxilonga ezijwayelekile ezisebenzisa i-common
noma izindawo zokusebenza ze-digital radiographic kanye ne-fluoroscopic
◆Izici zokufakwa: 12.5° Rhenium-Tungsten molybdenum target (RTM)
◆Izindawo ezigxile: Ezincane 0.6, Ezinkulu: 1.2
◆I-voltage enkulu yeshubhu: 125kV
◆Ihambisana nezimbozo zekhebula ezinamandla aphezulu zohlobo lwe-IEC60526
◆Ijeneretha ye-voltage ephezulu kufanele ivumelane ne-IEC60601-2-7
◆Ukuhlukaniswa kwe-IEC (IEC 60601-1:2005):IMISEBENZI YE-Class I ME

Imininingwane Yomkhiqizo

Imigomo Yokukhokha Nokuthumela:

Amathegi Omkhiqizo

Ireferensi Ejwayelekile

Ireferensi Ejwayelekile

Izihloko

I-EN 60601-2-54:2009 Imishini kagesi yezokwelapha - Ingxenye 2-54: Izidingo ezikhethekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile kwemishini ye-X-ray ye-radiography kanye ne-radioscopy
IEC60526 Ipulaki yekhebula ene-voltage ephezulu kanye nokuxhumeka kwesokhethi kwemishini ye-X-ray yezokwelapha
I-IEC 60522:1999 Ukunqunywa kokuhlunga okungapheli kwemihlangano ye-X-ray tube
I-IEC 60613-2010 Izici zikagesi, ukushisa kanye nokulayisha kwamashubhu e-X-ray anode ajikelezayo ukuze kutholakale ukuxilongwa kwezokwelapha
IEC60601-1:2006 Imishini kagesi yezokwelapha - Ingxenye 1: Izidingo ezijwayelekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile
I-IEC 60601-1-3:2008 Imishini kagesi yezokwelapha - Ingxenye 1-3: Izidingo ezijwayelekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile - Izinga Lokubambisana: Ukuvikelwa kwemisebe emishinini ye-X-ray yokuxilonga
IEC60601-2-28:2010 Imishini kagesi yezokwelapha - Ingxenye 2-28: Izidingo ezikhethekile zokuphepha okuyisisekelo kanye nokusebenza okubalulekile kwemihlangano yamashubhu e-X-ray yokuxilongwa kwezokwelapha
I-IEC 60336-2005 Imishini kagesi yezokwelapha-imihlangano yamashubhu e-X-ray yokuxilongwa kwezokwelapha-Izimpawu zezindawo ezigxile kuzo

Incazelo

●Igama lakhiwe kanje:

MWHX7110A

Ipayipi

A

Isokhethi ye-voltage ephezulu enesiqondiso sama-degree angu-90

MWTX71-0.6/1.2-125

B

Isokhethi ye-voltage ephezulu enesiqondiso sama-degree angu-270

Idatha Yobuchwepheshe

Impahla

Imininingwane

Okujwayelekile

Amandla okufaka ajwayelekile e-anode

F 1

F 2

I-IEC 60613

20kW(50/60Hz)

40kW(50/60Hz)

 

Umthamo wokugcina ukushisa we-Anode

110 kJ (150kHU)

I-IEC 60613

Umthamo omkhulu wokupholisa we-anode

500W

 
Umthamo wokugcina ukushisa

900kJ

 
Ukushisa okuqhubekayo okuphezulu ngaphandle komoya ojikelezayo

180W

 
Izinto ze-anodeIzinto zokumboza eziphezulu ze-Anode

I-Rhenium-Tungsten-TZM(RTM)

I-Rhenium-Tungsten-(RT)

 
I-engeli eqondiwe (Inkomba: i-axis yokubhekisela)

12.5 °

I-IEC 60788

Ukuhlanganiswa kwe-X-ray tube filtration engokwemvelo

1.5 mm Al / 75kV

I-IEC 60601-1-3

Inani(amanani) lendawo yokugxila

F1 (ukugxila okuncane)

F2 (ukugxila okukhulu)

I-IEC 60336

0.6

1.2

 
I-voltage ejwayelekile ye-X-ray tubeI-X-rayI-Fluoroscopic

125kV

100kV

I-IEC 60613

Idatha ngokushisa kwe-cathode Umkhawulo wamanje

I-voltage ephezulu

≈ /AC, < 20 kHz

 

F1

F 2

 

5.1A

≈79V

5.1 A

≈1214 V

 
Imisebe evuzayo ku-150 kV / 3mA ngebanga elingu-1m

0.5mGy/h

IEC60601-1-3

Insimu yemisebe ephezulu

443×443mm ku-SID 1m

 
Isisindo somhlangano we-X-ray tube

Cishe 18 kg

 

Izimo zokusebenza, ukugcina kanye nokuthuthwa

Imikhawulo

Imikhawulo Yokusebenza

Imikhawulo Yezokuthutha Neyokugcina

Izinga lokushisa elizungezile

Kusukela ku-10kuya ku-40

Kusukela ku- 20to 70

Umswakama ohlobene

≤75%

≤93%

Ingcindezi ye-barometric

Kusukela ku-70kPa kuya ku-106kPa

Kusukela ku-70kPa kuya ku-106kPa

 

Amanani okhiye be-Stator

i-stator yesigaba esi-1

Iphuzu lokuhlola

C-M

C-A

Ukumelana nokugoba

≈18.0…22.0Ω

≈45.0…55.0Ω

I-voltage yokusebenza evumelekile ephezulu (ukusebenza)

230V±10%

Ncoma i-voltage yokusebenza (ukusebenza)

160V±10%

I-voltage yokubhuleka

70VDC

I-voltage esebenzayo ekuchayekeni

Ama-Vrm angu-80

I-voltage egijimayo ku-fluoroscopy

Ama-VV angu-20-40

Isikhathi sokusebenza (kuye ngokuthi uhlelo lokuqalisa lunjani)

1.2s

Isexwayiso Ekusebenzeni

1. Imisebe ye-X-rayisivikelo

Lo mkhiqizo uhlangabezana nezidingo ze-IEC 60601-1-3.

Lo mhlangano we-X-ray tube ukhipha imisebe ye-X-ray isebenza. Ngakho-ke ngabasebenzi abafanelekayo nabaqeqeshwe ngokufanayo kuphela abavunyelwe ukusebenzisa umhlangano we-X-ray tube.

Imiphumela efanele yomzimba ingabangela ukulimala kwesiguli, abakhiqizi besistimu kufanele bavikeleke kahle ukuze bagweme imisebe ye-ionization.

2. I-Dielectric 0il

Ukuhlanganiswa kweshubhu le-X-ray kune-0il equkethwe yi-dielectric ukuze kuqiniswe ugesi ophezulu. Njengoba kuyingozi empilweni yabantuuma ivezwe endaweni engavinjelwekufanele ilahlwe ngokulandela umthetho wendawo.

3.Umkhathi Wokusebenza

Ukuhlanganiswa kweshubhu le-X-ray akuvunyelwe ukusetshenziswa emoyeni wegesi evuthayo noma ebolile ·

4Lungisa i-Tube Current

Kuye ngezimo zokusebenzaizici ze-filament zingase zishintshe.

Lolu shintsho lungase luholele ekuchayekeni ngokweqile kwe-X-ray tube assembly.

Ukuvimbela ukuthi i-X-ray tube assembly ingonakalilungisa njalo ugesi wepayipi.

Ngaphandle kwalokho lapho ithubhu le-X-ray linenkinga yokugobalukusetshenziswa kwesikhathi esideKudingeka ukulungiswa komsinga wepayipi.

5Izinga lokushisa lezindlu ze-X-ray Tube

Ungathinti indawo engaphansi kwe-X-ray tube ngemuva nje kokuhlinzwa ngenxa yokushisa okuphezulu.

Ishubhu le-X-ray elihlala lipholile.

6.Imikhawulo yokusebenza

Ngaphambi kokusebenzisasicela uqinisekise ukuthi isimo semvelo singaphakathi kwe-Iimits yokusebenza.

7Noma yikuphi ukungasebenzi kahle

P1lula ukuxhumana ne-SAILRAY ngokusheshauma kukhona ukungasebenzi kahle kwe-X-ray tube assembly okubonakalayo.

8. Ukulahlwa

Ukuhlanganiswa kweshubhu le-X-ray kanye neshubhu kuqukethe izinto ezifana namafutha nezinsimbi ezisindayo okumele kuqinisekiswe ukuthi ukulahlwa kwazo okungenangozi emvelweni nokufanelekile ngokuhambisana nemithethonqubo yezomthetho kazwelonke esebenzayo kumele kuqinisekiswe. Ukulahlwa kukadoti wasekhaya noma wezimboni akuvunyelwe. Umenzi unolwazi lobuchwepheshe oludingekayo futhi uzobuyisela ukuhlanganiswa kweshubhu le-X-ray ukuze kulahlwe.

Sicela uxhumane nesevisi yamakhasimende ngale njongo.

Amajika okukhishwa kwe-cathode

图片

Uma(A) Indawo Encane Yokugxila

Ikhasi 1.2

Uma(A) Indawo Enkulu Yokugxila

Ukulayisha okukodwa kanye nochungechunge

Izimo: I-Tube Voltage Isigaba Esithathu

Imvamisa Yamandla e-Stator 50Hz/60Hz

Ngiyi)

图片4

ama-t(s)

Ngiyi)

图片3

ama-t(s)

Ijika lokushisa nokupholisa le-anode

  IEC60613

图片5

Ukushisa nokupholisa ijika lomhlangano weshubhu le-X-ray

Izici Zokushisa Kwezindlu

图片6

Imidwebo yobukhulu be-X-ray tube assembly

SRMWHX7110A

图片7

Ukuhlanganiswa Kwesihlungi Nesigaba Esinqamulayo Sechweba

Isiqephu 27

Izintambo Zesixhumi Se-Rotor

Isiqephu 28

  • Okwedlule:
  • Olandelayo:

  • Inani Elincane Le-oda: 1pc

    Intengo: Ingxoxo

    Imininingwane Yokupakisha: 100pcs ngebhokisi ngalinye noma ngezifiso ngokuya ngobuningi

    Isikhathi Sokulethwa: 1 ~ 2 amasonto ngokuya ngobuningi

    Imigomo Yokukhokha: 100% T/T kusengaphambili noma i-WESTERN UNION

    Amandla Okunikezela: 1000pcs/ngenyanga

    Bhala umlayezo wakho lapha bese uwuthumela kithi